The Puritan Bennett 7200 respiratory ventilator has an advanced micro-processor design that offers CMV, assist/control, SIMV, CPAP, PEEP, and sigh controls. This critical care ventilator has a built-in nebulizer, air/oxygen blender, and variable pressure waveforms. Data is clear on the digital display and easy to input with the keyboard control.
Product Features
State-of-the-art gas delivery system.
Demonstrated track record of reliability
Full-featured microprocessor-controlled ventilator with 7202 display screen.
Gas delivery features - Flow-by®, Pressure Support, Pressure Control Ventilation
Patient monitoring features - Graphics 2.0, Respiratory Mechanics, Auto-PEEP
Communications features - RS-232, Digital Communications Interface (DCI 2.0)
Optional 7250® Metabolic Monitor
Integrated safety systems
Optional humidifier is available with this unit.
Indications For Usage For use in providing pediatric and adult patients with ventilatory support in the hospital environment.
Specifications:
Ventilator Dimensions: 41.9 cm H x 57.2 cm D x 55.9 cm W (16.5 in x 22.5 in x 22 in
Ventilator Weight: 50.8 kg (112 lbs)
Operator Selected Parameters Tidal Volume: .1 to 2.5 liters Respiratory Rate: .5 to 70 bpm Max Peak Inspiratory Flow: 120 Ipm during mandatory breather; 180 Ipm during spontaneous breathing. Sensitivity Inspiratory: .5 to 20 cmH2O bellow PEEP Oxygen Percentage: 21% to 100% Plateau: 0 to 2 seconds PEEP / CPAP: .1 to 2.5 liters, not to exceed twice tidal volume Sigh Rate and Frequency: 1 to 15 sighs per hour; 1 to 3 sighs per event.
Operator Selected Alarm Thresholds High Airway Pressure: 10 to 120 cmH2O Low Airway Pressure: 3 to 99 cmH2O PEEP / CPAP Pressure: 0 to 45 cmH2O Exhaled Tidal Volume: 0 to 2.5 liters Low Exhaled Minute Volume: 0 to 60 liters High Respiratory Rate: 0 to 70 bpm Sigh High Airway Pressure: 10 to 120 cmH2O, not to exceed twice high airway pressure. Ventilatory Modes: Operator selectable. Continuous Manditory Ventilation (CMV). Synchronous Intermittent Mandatory Ventilation (SIMV). Continuous Positive airway Pressure (CPAP). Inspiratory Flow Waveforms: Square, descending ramp, or sine. Alarm Control Functions: Alarm silence: silences alarm for 2 minutes. Alarm reset: resets ventilator to pre-alarm state of alert. Parameters Display: Airway pressure; exhaled volume; type of breath: assist, spontaneous, sigh, and plateau; mean airway pressure; peak airway pressure; PEEP/CPAP pressure; plateau pressure; respiratory rate; I:E ratio; tidal volume; minute volume; and spontaneous minute volume.
The Puritan Bennett® 840™Ventilator System is the flagship product in our line of critical care ventilators. It is responsive to patients and offers superior comfort, delivering...
Bear Cub® 750psv Infant Ventilator is designed to accommodate a range of patients from the small neonate to the 30kg pediatric patient. The 750psv incorporates Integrated Synchronized...