Bird V.I.P Standard Infant and Pediatric Ventilator
Call
- Brand
- Viasys
- SKU:
- TD VIP-REB
- Condition:
- Refurbished
- Weight:
- 0.00 LBS
This microprocessor VIP ventilator for infants and pediatrics provides assist/control and SIMV/CPAP in both volume cycled and time cycled modes. The Bird V.I.P. is responsiveness breath by breath, in an infant-pediatric ventilator, providing information that is accurate, responsive and excels to the highest levels of performance. The V.I.P. Bird® Ventilator System was designed specifically for neonatal, infant, and pediatric populations. It incorporates highly sophisticated features such as Flow Synchronization, Termination Sensitivity and a Dual Mode Breath Type to promote patient synchrony and facilitate weaning. The V.I.P. Bird® Ventilator system sets the premium standard for superior respiratory care in the NICU and PICU environments. The V.I.P. Bird® Ventilator System reaches the highest levels of performance by responding to a patient's ventilatory requirements breath by breath. This is achieved by utilizing software technology that continually evaluates the patients' breathing, allowing interaction between the infant and the ventilator while while providing the clinician with information that is accurate and responsive. Total Flow Synchrony - to match the patient's respiratory pattern. Pressure Support - available in all breath types Broad range for volume delivery 10 - 1200 ml "Smart" Sensor technology - autoclavable proximal flow sensor.
CONTROLS |
Power | On/Off (I/O) | ||
AC/DC Switch | AC Line/12 VDC | ||
Modes: | |||
vviiTime Cycled | Assist/Control, (S)IMV/CPAP | ||
vviiVolume Cycled | Assist/Control, SIMV/CPAP | ||
vviiPressure Support | Active in Volume SIMV/CPAP | ||
Tidal Volume | 20 to 995 ml | ||
Inspiratory Time | 0.10 to 3.00 sec. | ||
Breath Rate | 0 to 150 bpm | ||
Flow: | |||
vviiTime Cycled | 3 to 40 lpm | ||
vviiVolume Cycled | 3 to 100 lpm | ||
High Pressure Limit: | |||
vviiTime Cycled | 3 to 80 cmH2O | ||
vviiVolume Cycled | 3 to 120 cmH2O | ||
PEEP/CPAP | 0 to 24 cmH2O, Off | ||
Assist Sensitivity: | |||
vviiFlow Triggered | 0.2 to 5.0 lpm, Off | ||
vviiPressure Triggered | 1 to 20 cmH2O, Off | ||
Pressure Support | 1 to 50 cmH2O, Off | ||
Termination Sensitivity | Time Cycled, Flow Synchronized 5% to 25% of Peak Flow, Off | ||
Manual Breath | x 1 | ||
Oxygen Percent | 21% to 100% | ||
Over-Pressure Relief | 0 to 130 ± 10 cmH2O | ||
MONITORS & INDICATORS | |||
|
| ||
ALARMS | |||
All alarms feature both audible and visual indicators | |||
High Pressure Limit | 3 to 120 cmH2O | ||
Low Peak Pressure | 3 to 120 cmH2O, Off | ||
Low PEEP/CPAP Pressure | -9 to + 24 cmH2O | ||
High/Prolonged Pressure: | |||
VVIITime Cycled | High Pressure Limit + 10 cmH2O, Auto set | ||
VVIIVolume Cycled | 3 to 120 cmH2O | ||
VVIIAll Modes VVIIVI(Prolonged Pressure) | PEEP/CPAP + 6 cmH2O | ||
Apnea Interval | 20, 40 and 60 sec. | ||
Low Inlet Gas Pressure | System Pressure | ||
Red Light Indicator Lamp | |||
Circuit Fault | Red Light Indicator Lamp | ||
PRE-SET VALUES | |||
Minimum Expiratory Time: | |||
VVIIAll Modes | 250 msec | ||
Minimum Inspiratory Time: | |||
VVIITime Cycled | 100 msec | ||
VVIIVolume Cycled | 150 msec | ||
Demand Flow: | |||
VVIISensitivity | 1 cmH2O below baseline pressure | ||
VVIIFlow Range | 0 - 120 lpm maximum | ||
Leak Compensation System: | |||
VVII(Volume Cycled) | |||
VVIIAssist Sensitivity 1: | Set at 1 cmH2O | ||
VVIIviActivation | 0.25 cmH2O below baseline pressure | ||
VVIIviFlow Range | 0 - 51 lpm | ||
VVIIAssist Sensitivity 2: | Set at 2 cmH2O | ||
VVIIviActivation | 0.50 cmH2O below baseline pressure | ||
VVIIviFlow Range | 0 - 10 lpm | ||
INPUTS | |||
Electrical | Fuse Ratings | ||
100 VAC+10% - 15%, 50/60 Hz | 0.75 amps 1.25 SB | ||
120 VAC+10% - 15%, 50/60 Hz | 0.58 amps 1.00 SB | ||
220 VAC+10% - 15%, 50/60 Hz | 0.35 amps 0.63 SB | ||
240 VAC+10% - 15%, 50/60 Hz | 0.32 amps 0.63 SB | ||
Pneumatic | |||
VVIIAir | 40 - 75 PSIG (2.1 - 4.9 kg/cm2) | ||
VVIIOxygen (O2) | 40 - 75 PSIG (2.1 - 4.9 kg/cm2) | ||
VVIIvi(Air and O2 must be within 20 PSIG) | |||
OUTPUTS | |||
Fiber Optic Data Link | PARTNER® Volume Monitor | ||
PARTNER® lli Volume Monitor | |||
Bird Graphics Monitor | |||
Auxiliary Gas | 0 to 15 lpm, FlO2 at blender setting |